InfoCons Consumer Protection Alert : TRANS MEDIKAL – GETINGE – Steam generator

TRANS MEDIKAL – GETINGE – Steam generator

Name: No Name Provided

Category: 10

Date: 2025-05-09

Product Category: Pressure equipment/vessels

Risk Type: Burns

Danger: Due to a manufacturing error, excess pressure can build up in the steam generator and a significant amount of energy could be released resulting in injuries, including burns. The product does not comply with the requirements of the Pressure Equipment Directive.

Measures: Type of economic operator to whom the measure(s) were ordered: DistributorCategory of measure(s): Recall of the product from end usersDate of entry into force: 02/03/2025Type of economic operator taking notified measure(s): DistributorCategory of measure(s): Making the marketing of the product subject to prior conditionsDate of entry into force: 02/03/2025

Description: Steam generator integrated into steriliser.

Notifying Country: France

Country of Origin: Türkiye

Alert Type: Professional

Alert Level: Other risk levels

TRANS MEDIKAL - GETINGE - Steam generatorTRANS MEDIKAL - GETINGE - Steam generator

Other blog posts