Pannonfitt – Laser therapy device
Name: Family Laser
Category: 10
Date: 2026-06-12
Product Category: Electrical appliances and equipment
Risk Type: Burns
Danger: The battery management system lacks a cell temperature monitoring device. The product can overheat and cause burns. The product does not comply with the requirements of the Low Voltage Directive nor with the European standards EN 62133 and EN 60335.
Measures: Type of economic operator to whom the measure(s) were ordered: ImporterCategory of measure(s): Withdrawal of the product from the marketDate of entry into force: 23/06/2025Type of economic operator to whom the measure(s) were ordered: ImporterCategory of measure(s): Recall of the product from end usersDate of entry into force: 23/06/2025
Description: Low-level laser therapy device for home use, including a USB charging cable and a pair of protective glasses. The handheld device is stored in a blue, zippered, hard-shell carrying case.
Notifying Country: Hungary
Country of Origin: People’s Republic of China
Alert Type: Consumer
Alert Level: Other risk levels